Z-0926-2013Class II

Medtronic O-arm Imaging System

Medtronic Navigation, Inc / initiated 2013-01-31 / Terminated / root cause: design

Medtronic Navigation, Inc. began a recall of Medtronic O-arm Imaging System Product Usage: image-intensified fluoroscopic x-ray system, mobile on . FDA classified it as Class II and gives the reason as "Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0926-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic O-arm Imaging System Product Usage: image-intensified fluoroscopic x-ray system, mobile
Product code
OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
Reason for recall
Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction
Root cause tag
design
FDA root cause
Component design/selection
Action
Medtronic Navigation issued a Field Safety Notice letter dated January 18, 2013 to all affected customers. The letter identified the affected product, problem, and actions to be taken. The letter instructed customers not to discontinue use and that a Medtronic Navigation Representative will be contacting them shortly to schedule a time to conduct the motion control box upgrade. For questions contact your local Medtronic Navigation representative or call Technical Services Group at 800-595-9709 or (720) 890-3200, or email at rs.navtechsupport@ medtronic.com.
Quantity in commerce
463 units
Distribution
Worldwide Distribution - USA Nationwide and the countries of: Canada, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CHILE, CHINA, CROATIA, CZECH REPUBLIC, DENMARK,FINLAND, FRANCE, GERMANY, GREECE, INDIA, IRELAND, ISRAEL, ITALY, JAPAN ,KOREA, REPUBLIC OF KUWAIT, LEBANON, LIBYAN ARAB JAMAHIRIYA, MEXICO, NETHERLANDS, New Zealand, POLAND, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, Taiwan(R.O.C.) , THAILAND, TURKEY, UNITED ARAB EMIRATES. and UNITED KINGDOM.
Product codes and lots
Product Catalog Numbers: 9732719, 9732719R, 9733346, 9733346R, BI-700-00027, BI-700-00027R, BI-700-00027-GER, BI-700-00027-GERR, BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R  Units with serial numbers 101 through 564 and unit 580 are affected. Serial numbers 565 and higher are not affected .

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0926-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0926-2013Class IITerminatedVoluntary: Firm initiated