Z-0896-2013Class II

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700)

Orthofix SRL / initiated 2013-02-01 / Terminated / root cause: design

Orthofix, Inc began a recall of Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod... on . FDA classified it as Class II and gives the reason as "There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0896-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.
Root cause tag
design
FDA root cause
Device Design
Action
Orthofix sent an Urgent Medical Device Recall Notification dated February 1, 2013 to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to identify and remove the affected products from inventory immediately and follow the actions listed in the enclosed Product Return Instructions. For questions or concerns contact Regulatory Affairs at 214-937-2100.
Quantity in commerce
565 total (287 PN 52-6800 and 278 PN 52- 6700)
Distribution
Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.
Product codes and lots
Catalog Numbers: 52-6800 & 52-6700; All Lot Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0896-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0896-2013Class IITerminatedVoluntary: Firm initiated