Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0812-2013Class III

Ultradent Products, Chromaclone PVS, Fast Set heavy body, Bubble Gum Flavor, 50 ml, Impression Material, Part...

Ultradent Products Inc. / initiated 2013-02-04 / Terminated / root cause: packaging

Ultradent Products, Inc. began a recall of Ultradent Products, Chromaclone PVS, Fast Set heavy body, Bubble Gum Flavor, 50 ml, Impression Material, Part No. 5013, Lot No. B7DS3, Used in prosthodontics on . FDA classified it as Class III and gives the reason as "Incorrect packaging of regular set and fast set cartridges". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0812-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ultradent Products Inc.
Product
Ultradent Products, Chromaclone PVS, Fast Set heavy body, Bubble Gum Flavor, 50 ml, Impression Material, Part No. 5013, Lot No. B7DS3, Used in prosthodontics.
Product code
ELW Material, Impression
Reason for recall
Incorrect packaging of regular set and fast set cartridges.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Ultradent Products, Inc. sent an Urgent Medical Device Correction letter dated February 4, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected products. They may continue to use product or dispose unwanted cartridges and reply back to firm indicating acknowledgement of letter and if replacement is needed. Consumers with questions can call the firm at 801-553-4420. For questions regarding this recall call 801-553-4420.
Quantity in commerce
100 cartridges
Distribution
Nationwide Distribution
Product codes and lots
Lot No. B7DS3

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0812-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0811-2013Class IIITerminatedVoluntary: Firm initiated
Z-0812-2013Class IIITerminatedVoluntary: Firm initiated