Z-1052-2013Class I

ShapeMatch Cutting Guides

Howmedica Osteonics Corp / initiated 2012-11-08 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of ShapeMatch Cutting Guides Product Usage: The ShapeMatch Cutting Guides are intended to be used as patient-specific surgical instrumentation to assist in the... on . FDA classified it as Class I and gives the reason as "With regards to ShapeMatch Cutting Guides using MRI Imaging, Stryker Orthopaedics has become aware that there are potential issues associated with internal processes for planning cases and producing ShapeMatch Cutting...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1052-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
ShapeMatch Cutting Guides Product Usage: The ShapeMatch Cutting Guides are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee arthroplasty components intraoperatively and in guiding the marking of bone before cutting.
Product code
MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Reason for recall
With regards to ShapeMatch Cutting Guides using MRI Imaging, Stryker Orthopaedics has become aware that there are potential issues associated with internal processes for planning cases and producing ShapeMatch Cutting Guides.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker sent Product Correction Bulletins on 11/8/2012 dated 11/9/2012 and 11/19/2012 to all affected customers. Stryker sent an updated Urgent Medical Device Product Notification letters and Acknowledgement letters dated 1/24/2013 through 1/28/2013 to all affected customers via Fed Ex with return receipt: The letter identified the affect product, problem, potential hazards, risk mitigating factors and actions to be taken. For questions call 201-831-6878. Stryker issued press release on 4/10/2013.
Quantity in commerce
7868 units
Distribution
USA Nationwide Distribution
Product codes and lots
Catalog No: TR7100-L and TR7100-R Lot Code: All

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1052-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1052-2013Class ITerminatedVoluntary: Firm initiated