Z-0920-2013Class II

Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA...

Integra LifeSciences Corporation / initiated 2013-01-07 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System... on . FDA classified it as Class II and gives the reason as "There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0920-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.
Product code
LBK Neurosurgical Ultrasonic Instruments
Reason for recall
There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. This erosion can potentially lead to tip breakage. There were no reports of a patient injury associated with these complaints.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Integra sent a Urgent Medical Device Recall Field Safety Notices dated January 7, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed of the nature of the issue and how if affects CUSA top erosion/breakage; advises customers and provided a copy of the CEM Nosecone IFU additional Warning; and requess customers to complete and return the Acknowledgement Return Form. Please feel free to contact our service hotline at ( 888 ) 772-7378.
Quantity in commerce
USA 6,190, EU 22,972 and ROW 288 units
Distribution
Worldwide Distribution - USA and the countries of Austria, Australia, Bangladesh, Belgium, Bulgaria, Brazil, Switzerland, Canada, China, Costa Rica, Czech Republic, Germany, Dominican Republic, Egypt, Emirates, Spain, Finland, France, Great Britian, Hong Kong, Hungary, Indonesia, Ireland, Israel, India, Italy, Japan, Korea (South), Kuwait, Lebanon, Sri Lanka, Latvia, Macedonia, Mexico, Malaysia, Netherlands, Philippines, Pakistan, Poland, Portugal, Qatar, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Tunisia, Taiwan, United Arab, Viet Nam and South Africa.
Product codes and lots
CEM Nosecone Catalogue No. C6623 Lot numbers: 1092855, 1093129, 1100436, 1100437, 1101217, 1101328, 1103393, 1104265, 1105067, 1110311, 1110892, 1112182, 1112183, 1114112, 1114655, 1115225, 1115499, 1115500,  1120771, 1121141, 1121142, 1121383, 1121627, 1122258, 1122259, 1123049 and 1130710.  CEM Nosecone, Catalogue No. C6636: 1091431, 1091679, 1091680, 1093475, 1095103, 1095416, 1095417, 1100242, 1100434, 1100435, 1101094, 1101218, 1101219, 1103104, 1103395, 1103849, 1103849, 1103850, 1103852, 1103917, 1103918, 1103922, 1104257, 1104263, 1105071, 1110475, 1110476, 1110978, 1111490, 1111493, 1112187, 1112188, 1112212, 1113179, 1113180, 1113663, 1114357, 1114358, 1114659, 1114660, 1115039, 1115502, 1115504, 1115505, 1120776, 1120777, 1121167, 1121168, 1121169, 1121565, 1121566, 1121567, 1122344, 1123051, 1123052, 1123508, 1123511 and 1124053.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0920-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0920-2013Class IITerminatedVoluntary: Firm initiated