Z-0841-2013Class II

Monaco RTP System. Radiation treatment planning

Elekta Limited / initiated 2013-02-11 / Terminated / root cause: manufacturing process

Elekta, Inc. began a recall of Monaco RTP System. Radiation treatment planning on . FDA classified it as Class II and gives the reason as "The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0841-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Monaco RTP System. Radiation treatment planning.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Elekta provided customers with an "Important Safety Notice - Dose Display Not Matching Beam Geometry - Monaco RFC 35748". The notice identified the product, problem, and actions to be taken by the customers. Contact the form at 800-878-4267 for questions relating to this notice.
Quantity in commerce
13
Distribution
Nationwide Distribution-including the states of AK, CA, FL, NJ, NC, OH, OR, TN, TX, WA, and WI.
Product codes and lots
Versions 3.20.00 and 3.20.01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0841-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0841-2013Class IITerminatedVoluntary: Firm initiated