Z-0841-2013Class II
Monaco RTP System. Radiation treatment planning
Elekta Limited / initiated 2013-02-11 / Terminated / root cause: manufacturing process
Elekta, Inc. began a recall of Monaco RTP System. Radiation treatment planning on . FDA classified it as Class II and gives the reason as "The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0841-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Monaco RTP System. Radiation treatment planning.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Elekta provided customers with an "Important Safety Notice - Dose Display Not Matching Beam Geometry - Monaco RFC 35748". The notice identified the product, problem, and actions to be taken by the customers. Contact the form at 800-878-4267 for questions relating to this notice.
- Quantity in commerce
- 13
- Distribution
- Nationwide Distribution-including the states of AK, CA, FL, NJ, NC, OH, OR, TN, TX, WA, and WI.
- Product codes and lots
- Versions 3.20.00 and 3.20.01
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0841-2013 | Class II | Terminated | Voluntary: Firm initiated |