INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4...

Smith & Nephew, Inc. / initiated 2013-02-07 / Terminated / root cause: labeling and UDI

Smith & Nephew Inc began a recall of INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 75 MM, TITANIUM, Smith & Nephew, Inc... on . FDA classified it as Class II and gives the reason as "75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0944-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 75 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for intramedullary rod fixation.
Product code
JDS Nail, Fixation, Bone
Reason for recall
75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Smith & Nephew sales representatives and Smith & Nephew international distributors were notified of problem and the recall via email on February 7, 2013. They were instructed to cease use and distribution of the product and quarantine for return. Non responding consignees were notified again on February 28, 2013 by email. Hospitals were notified by letter beginning on February 7, 2013. Smith & Nephew also contacted the surgeons who implanted the screws by letter the week of February 11th. For questions regarding this recall call 978-749-1330.
Quantity in commerce
90 units
Distribution
Worldwide Distribution - USA including NM, TN, TX, NC, CA, MI, KY, MA, NJ, OH, and HI. Internationally to Canada, United Arab Emirates, Mexico, Argentina, Japan, and Chile.
Product codes and lots
Batch Number 12JM17051

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0944-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0943-2013Class IITerminatedVoluntary: Firm initiated
Z-0944-2013Class IITerminatedVoluntary: Firm initiated