Z-0925-2013Class II

REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument

Zimmer GmbH / initiated 2013-02-05 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument on . FDA classified it as Class II and gives the reason as "The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0925-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
BIOMET sent an Urgent Recall Notice beginning February 13, 2013, to all affected customers. The notice included possible adverse events and directions to locate and remove identified devices by returning a Fax Back Response Form to 574-372-1683 and obtaining an RGA # prior to return of product to Biomet, Inc. 56 East Bell Drive Warsaw, IN 46580. Questions should be directed to 574 372-1570 M-F, 8 a.m.-5.p.m. For questions regarding this recall call 574-267-6639, ext 1676.
Quantity in commerce
58
Distribution
Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile
Product codes and lots
Catalog Number 5313152 Lot 153880,153930, 415500, 648010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0925-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0925-2013Class IITerminatedVoluntary: Firm initiated