Z-0948-2013Class II
Siemens Somatom CTs equipped with software syngo CT2010A, syngo CT2010B, syngo CT2011A and CT2012B. Computed...
Siemens Medical Solutions USA, Inc. / initiated 2012-11-27 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Siemens Somatom CTs equipped with software syngo CT2010A, syngo CT2010B, syngo CT2011A and CT2012B. Computed tomography x-ray system on . FDA classified it as Class II and gives the reason as "Sporadic hang-up malfunctions concerning 3D reconstructions during activated "Preview Image" functionality". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0948-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Somatom CTs equipped with software syngo CT2010A, syngo CT2010B, syngo CT2011A and CT2012B. Computed tomography x-ray system.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Sporadic hang-up malfunctions concerning 3D reconstructions during activated "Preview Image" functionality.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" dated November 27, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The correction for this problem will be available with Service Pack 1 for syngo CT2011A And Service Pack 4 for syngo CT2012B. Both Service pack are planned to be released starting February 2013. Systems still running on syngo CT2010A or syngo CT2010B are already scheduled to be upgraded to syngo CT2011A or syngo CT2012B respectively. Contact your local Siemens Application Specialist for additional information.
- Quantity in commerce
- 896
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Model numbers - 7740769, 8098027, 10430603, and 10590000.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0948-2013 | Class II | Terminated | Voluntary: Firm initiated |