Z-1323-2013Class II

Heartware HVAD Pump Implant Kit (Heartware Ventricular Assist Device)

HeartWare, Inc. / initiated 2012-12-27 / Terminated / root cause: design

Heartware Inc began a recall of Heartware HVAD Pump Implant Kit (Heartware Ventricular Assist Device) on . FDA classified it as Class II and gives the reason as "HeartWare, Inc. of Miami Lakes, FL is recalling their Ventricular Assist System due to the loosening of the HVAD Pump driveline connector". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1323-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
HeartWare, Inc.
Product
Heartware HVAD Pump Implant Kit (Heartware Ventricular Assist Device)
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
HeartWare, Inc. of Miami Lakes, FL is recalling their Ventricular Assist System due to the loosening of the HVAD Pump driveline connector.
Root cause tag
design
FDA root cause
Device Design
Action
HeartWare, Inc. sent an Urgent Medical Device Correction letter to all their customers on February 7, 2013 instructing that at each routine clinic visit, they should inspect the patient's driveline connector as described in the attached Technical Bulletin. They were asked to sign and return the acknowledgement form to HeartWare. For questions contact your HeartWare representative or Customer Service at 1-877-367-4823.
Quantity in commerce
3,544
Distribution
Worldwide Distribution-USA (nationwide) including the states of MN, IL, MI, FL, WI, TX, MO, CA, MA, TN, OH, NC, GA, VA, UT, KY, MD, IN, PA, NY, NJ, LA, OR, WA, AL and CO, and the countries of Poland, Australia, Austria, Italy, Germany, New Zealand, South Africa, Czech Republic, France, Brazil, Switzerland, Singapore, United Kingdom, Saudi Arabia, Netherlands, China, Chile, Belgium, Canada, Norway, Lebanon, Israel, Malaysia, Turkey, India, Argentina, Greece, Japan and Lithuania.
Product codes and lots
Serial Numbers: United States: HW001-HW3827 and International: HW001-HW3827 and HW1000-HW10675