Z-0991-2013Class II

VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100...

bioMerieux, Inc. / initiated 2013-02-18 / Terminated / root cause: software

Biomerieux Inc began a recall of VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100 8-port, Cat. #W0441; (2) dc7100 8-port (refurbished)... on . FDA classified it as Class II and gives the reason as "There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0991-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100 8-port, Cat. #W0441; (2) dc7100 8-port (refurbished), Cat. #W0441R; (3) dc7700 8-port, Cat. #W0447; (4) dc7700 8-port (refurbished), Cat. #W0447R; (5) dc7800, Cat. #W0449; (6) dc7800 (refurbished), Cat. #W0449R (7) rp5700, Cat. #W0452; (8) rp5700 (refurbished), Cat. #W0452R; (9) rp5800 XPE, Cat. 413642; (10) rp5800WES7, Cat. #413862; (11) d530 6-port, Cat. #W0438; (12) d530 6-port (refurbished), Cat. #W0438R; and (13) d530 4-port, Cat. #W0437. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
Product code
LTW Susceptibility Test Cards, Antimicrobial
Reason for recall
There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.
Root cause tag
software
FDA root cause
Software design
Action
bioMerieux sent an "URGENT PRODUCT CORRECTION NOTICE" dated February 20, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Emails were issued by the recalling firm on 2/18/2013 and 2/19/2013 to their subsidiaries and to distributors. Letters began issuing to customers via e-mail on 2/20/2013. Contact bioMerieux Clinical Customer Service at (800) 682-2666 for questions regarding this notice.
Quantity in commerce
For all 3 computers types: 16,736 originally distributed; 554 refurbished and redistributed
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Algeria, Angola, Argentina, Australia, Austria, Azerbaidjian, Bahrain, Belarus, Belgium, Benin, Bolivia, Bosnia, Brazil, Bulgaria, Burkina Faso, Camaroun, Canada, Chad, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Democratic Republic of Congo, Demark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Guinea, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Netherlands, Netherlands, Nicaragua, Niger Niamey, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Yemen.
Product codes and lots
Manufactured from September, 2004, to present.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0991-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0989-2013Class IITerminatedVoluntary: Firm initiated
Z-0990-2013Class IITerminatedVoluntary: Firm initiated
Z-0991-2013Class IITerminatedVoluntary: Firm initiated