Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm
Covidien LP began a recall of Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm Product Code: 030450 Applications in abdominal, gynecologic, pediatric anti thoracic surgery... on . FDA classified it as Class II and gives the reason as "Single Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1063-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm Product Code: 030450 Applications in abdominal, gynecologic, pediatric anti thoracic surgery for resection, transection and creation of anastomosis.
- Product code
- GDW Staple, Implantable
- Reason for recall
- Single Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Covidien notified customers with an Urgent Medical Device Recall notification on February 11, 2013. The letter identified the affected product and the reason for the recall. Accounts were requested to quarantine and discontinue use of the affected devices, and notify Universal Loading Unit 30mm-2.0mm users of this recall. Customers should follow the instructions provided in order to return the affected product. Customers are to complete and return the Product Return Form regardless if they have any affected product or not. For questions regarding the RGA I return process, please contact Covidien Customer Service, M-F, 8 am - 6:30 pm ET at (800) 962-9888, option 1, and then option 2.
- Quantity in commerce
- 6,751 units
- Distribution
- Worldwide Distribution -- USA, Asia Pacific, Canada, Europe, and Japan.
- Product codes and lots
- Lot Codes: N2F0563LX, N2F0341LX, N2F0725LX, and N2E0605LX.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1063-2013 | Class II | Terminated | Voluntary: Firm initiated |