Z-1063-2013Class II

Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm

Medtronic, Inc. / initiated 2013-02-11 / Terminated / root cause: manufacturing process

Covidien LP began a recall of Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm Product Code: 030450 Applications in abdominal, gynecologic, pediatric anti thoracic surgery... on . FDA classified it as Class II and gives the reason as "Single Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1063-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm Product Code: 030450 Applications in abdominal, gynecologic, pediatric anti thoracic surgery for resection, transection and creation of anastomosis.
Product code
GDW Staple, Implantable
Reason for recall
Single Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Covidien notified customers with an Urgent Medical Device Recall notification on February 11, 2013. The letter identified the affected product and the reason for the recall. Accounts were requested to quarantine and discontinue use of the affected devices, and notify Universal Loading Unit 30mm-2.0mm users of this recall. Customers should follow the instructions provided in order to return the affected product. Customers are to complete and return the Product Return Form regardless if they have any affected product or not. For questions regarding the RGA I return process, please contact Covidien Customer Service, M-F, 8 am - 6:30 pm ET at (800) 962-9888, option 1, and then option 2.
Quantity in commerce
6,751 units
Distribution
Worldwide Distribution -- USA, Asia Pacific, Canada, Europe, and Japan.
Product codes and lots
Lot Codes: N2F0563LX, N2F0341LX, N2F0725LX, and N2E0605LX.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1063-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1063-2013Class IITerminatedVoluntary: Firm initiated