Hickman 9.0 French Dual Lumen Catheter with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous...
Bard Access Systems began a recall of Hickman 9.0 French Dual Lumen Catheter with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer system (pediatric product code 0600580; adult... on . FDA classified it as Class II and gives the reason as "Bard Access Systems is recalling two lots of Hickman 9.0 French and one lot of Leonard 10.0 French Dual Lumen Catheters with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer Systems due to the wrong...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0909-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- Hickman 9.0 French Dual Lumen Catheter with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer system (pediatric product code 0600580; adult product code 0600600). Product Usage: Hickman and Leonard Catheters are designed for long-term vascular access and for use in patients that lack adequate peripheral venous access.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Bard Access Systems is recalling two lots of Hickman 9.0 French and one lot of Leonard 10.0 French Dual Lumen Catheters with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer Systems due to the wrong size introducer being included with the kit.
- FDA root cause
- Mixed-up of materials/components
- Action
- Bard sent an Urgent Recall Notification letters dated February 2013 to all affected customers. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to examine their inventory for the affected product, return any remaining in stock and complete the enclosed Reply Form and Inventory Reconciliation Form. For questions contact Bard Access Systems Customer Service at 1-800-290-1689.
- Quantity in commerce
- 195 units.
- Distribution
- USA Nationwide Distribution and the countries Europe, Belgium, Australia, and Latin America.
- Product codes and lots
- Lot numbers: HUWI1379 and HUWI1380.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0909-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0910-2013 | Class II | Terminated | Voluntary: Firm initiated |