Z-0915-2013Class II

AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used...

STERIS Corporation / initiated 2013-02-12 / Terminated / root cause: software

Steris Corporation began a recall of AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities on . FDA classified it as Class II and gives the reason as "The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0915-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
Product code
FLE Sterilizer, Steam
Reason for recall
The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
Root cause tag
software
FDA root cause
Software design
Action
Steris Corporation initiated their recall on February 12, 2013. They will notify customers via onsite visits by STERIS Field Service Technicians to each affected Customer site. For further questions, please call (440) 392-7601.
Quantity in commerce
395 Units
Distribution
Worldwide Distribution - USA including Canada and the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, SK, TN, TX, UT, VA, VT, WA, WI, and WV.
Product codes and lots
AMSCO C: Serial Numbers: 021911201 023461202 AMSCO 400: Serial Numbers: 030181215-030291310

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0915-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0915-2013Class IITerminatedVoluntary: Firm initiated