Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0396-2014Class II

Synthes Titanium Minimally Invasive Reduction Screw System Locking Cap, Part Number 04.631.502 Non-cervical...

Synthes GmbH / initiated 2012-09-07 / Terminated

Synthes USA HQ, Inc. began a recall of Synthes Titanium Minimally Invasive Reduction Screw System Locking Cap, Part Number 04.631.502 Non-cervical spinal fixation devices intended for posterior... on . FDA classified it as Class II and gives the reason as "This action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0396-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Titanium Minimally Invasive Reduction Screw System Locking Cap, Part Number 04.631.502 Non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
This action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final tightening, the potential exists for locking caps to loosen post-operatively in certain situations.
FDA root cause
Use error
Action
Synthes sent an Urgent: Medical Device Product Removal letter on September 7, 2012, to all affected consignees. Consignees were requested to immediately cease using the product, examine their inventory and remove the parts from inventory. If they had the parts, they were further instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number, complete the verification section at the end of the letter and return the verification form. If they did not have the parts, they were requested to complete the verification section at the end of the letter and return the verification form. Consignees with questions were instructed to call 610-719-5450. For questions regarding this recall call 610-719-5000.
Quantity in commerce
745
Distribution
Nationwide Distribution including CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0396-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0395-2014Class IITerminatedVoluntary: Firm initiated
Z-0396-2014Class IITerminatedVoluntary: Firm initiated