Z-0939-2013Class II

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual...

Draeger, Inc. / initiated 2012-12-13 / Terminated / root cause: labeling and UDI

Draeger Medical, Inc. began a recall of Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system on . FDA classified it as Class II and gives the reason as "The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0939-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Drager sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-543-5047 for questions relating to this notice.
Quantity in commerce
260
Distribution
Distributed in the states of CA, IL, and MA.
Product codes and lots
Catalog Number MS20783 (Infinity MCable-Dual Hemo)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0939-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0939-2013Class IITerminatedVoluntary: Firm initiated