Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1028-2013Class II

Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the...

Roche Diagnostics GmbH / initiated 2013-03-01 / Terminated / root cause: manufacturing process

Roche Diagnostics Operations, Inc. began a recall of Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas CT/NG 4800 System, for in vitro diagnostics on . FDA classified it as Class II and gives the reason as "Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1028-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas CT/NG 4800 System, for in vitro diagnostics.
Product code
OOI Real Time Nucleic Acid Amplification System
Reason for recall
Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Roche sent a Urgent Medical Device Correction letter via UPS to all affected customers ( receipt required) The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to visually inspect affected product for the molding defect and to verify equal distribution of reagent in both chambers of the reservoir after adding reagent. All customers are requested to fill out the FAX Back form to schedule shipment of defective part. Non-responding accounts will be monitored on an ongoing basis and follow-up attempts will be made. Questions and concerns are to be addressed to Roche Molecular Diagnostics Technical Support 24 hours/day 1-800-526-1247.
Quantity in commerce
189 units
Distribution
Nationwide Distribution including Puerto Rico and the states of MO, NE, CO, MT, NJ, FL, WA, TX, GA, NY, KS, NH, VA, HI, PA, MI, IA, CA, NC, AZ, UT, LA, IL, TN, ME, MN, KY, CT, OR, NV, AL and NM.
Product codes and lots
Lot / SN : 477415,477416, 477898, and 477899

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1028-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1028-2013Class IITerminatedVoluntary: Firm initiated