Z-1049-2013Class II

Siemens Dimension(R) Phosphorus (PHOS) Flex(R) reagent cartridge (DF61) The Phosphorus (PHOS) method used on...

Siemens Healthcare Diagnostics Inc. / initiated 2013-02-22 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension(R) Phosphorus (PHOS) Flex(R) reagent cartridge (DF61) The Phosphorus (PHOS) method used on the Dimension(R) clinical chemistry system is an... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has initiated a voluntary field action for Dimension(R) PHOS Flex(R) Reagent due to reagent interaction causing falsely depressed Creatinine results when EZCR is processed from open wells...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1049-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Dimension(R) Phosphorus (PHOS) Flex(R) reagent cartridge (DF61) The Phosphorus (PHOS) method used on the Dimension(R) clinical chemistry system is an in vitro diagnostic test intended for the quantitative measurement of phosphorus in human serum, plasma, and urine.
Product code
CEO Phosphomolybdate (Colorimetric), Inorganic Phosphorus
Reason for recall
Siemens Healthcare Diagnostics has initiated a voluntary field action for Dimension(R) PHOS Flex(R) Reagent due to reagent interaction causing falsely depressed Creatinine results when EZCR is processed from open wells that are in close proximity to open wells of PHOS reagent.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent an Urgent Medical Device Correction letter dated February 22 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken by Customer: 1. Customers with two or more Dimension systems should perform testing of PHOS and EZCR on separate Dimension systems. 2. Customers with only one Dimension system should discontinue use of either PHOS or EZCR on the Dimension system. Please contact your local sales representative to discuss alternate solutions that will best meet the needs of your laboratory. Please complete the attached Effectiveness Check form and fax it to the Siemens Technical Solutions Center at (302) 631-8467, to indicate you have received the information. If you have any questions, please contact the Siemens Technical Solutions enter or your local Siemens technical support representative. We apologize for the disruption this situation will cause your laboratory. We are working diligently to resolve this issue with a permanent solution. .
Quantity in commerce
60,674
Distribution
Nationwide Distribution including the states of AK, AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
Product codes and lots
Lot numbers DC3078, exp 3/19/13; DA3092, exp 4/2/13; FB3109, exp 4/19/13; BC3127, exp 5/7/13; BB3152, exp 6/1/13; EC3166, exp 6/15/13; BB3190, exp 7/9/13; DA3212, exp 7/31/13; EA3233, exp 8/21/13; FB3260, exp 9/17/13; DA3319, exp 11/15/13; DA3320, exp 11/16/13; GC3337, exp 12/3/13; FB3346, exp 12/12/13; CB4014, exp 1/14/14; and FB4029, exp 1/29/14.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1049-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1049-2013Class IITerminatedVoluntary: Firm initiated