Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0959-2013Class II

Allen Traction Boot, Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders...

Allen Medical Systems, Inc. / initiated 2013-02-28 / Terminated / root cause: manufacturing process

Allen Medical Systems, Inc. began a recall of Allen Traction Boot, Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders, non-invasive medical devices designed to attach to... on . FDA classified it as Class II and gives the reason as "Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0959-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Allen Medical Systems, Inc.
Product
Allen Traction Boot, Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders, non-invasive medical devices designed to attach to orthopedic fracture tables to aid clinical staff in applying traction to the leg of a patient during surgical procedures.
Product code
HST Apparatus, Traction, Non-Powered
Reason for recall
Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Allen Medical Systems sent a "Urgent Medical Device Recall" letter dated February 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken. Customers were instructed to: 1. Immediately remove from service your Allen Traction Boot with the serial number(s) XXXXX 2. Contact Allen Medical Customer Service at (800) 433-5774 for an RMA number and instructions to return your Allen Traction Boot for inspection and repair if necessary. There will be no cost to you. 3. Return your Allen Traction Boot to Allen Medical as soon as possible. In order to minimize the inconvenience, we will repair your Allen Traction Boot and return the device to you within 48 hours of receipt.
Quantity in commerce
24 devices
Distribution
Worldwide Distribution- USA including the states of TX, NV, KY, GA, OH, PA, NY, AL, WA, IL and the countries of New Zealand and Japan.
Product codes and lots
A260264, A253312, A260989, A260262, 418292, A260991, A260998, A253313, A260993, A260265, 418293, A253309, A260995, A253310, A260992, A253311, 418294, A253323, A260261, A253321, A260994, 418291, 418295, A253322

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0959-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0959-2013Class IITerminatedVoluntary: Firm initiated