Arrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 30 cc; Product Number...
Arrow International Inc began a recall of Arrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 30 cc; Product Number: IAB-05830-LWS. The IAB is utilized for intra-aortic balloon... on . FDA classified it as Class II and gives the reason as "Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1064-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 30 cc; Product Number: IAB-05830-LWS. The IAB is utilized for intra-aortic balloon counter-pulsation therapy.
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Arrow International issued Urgent Medical Device Recall notification on 11/27/12. The letter identified the affected product and the reason for the recall. Accounts are requested to check their stock for the products included within the scope of this recall. Also, customers should cease use and distribution, and quarantine all affected product immediately. Contact Arrow's Customer Service Department at 800-523-8446 for a Return Authorization Number. Customers are to complete and fax the Recall Acknowledgement & Stock Status Form to the number provided, and return the affected product. If customers have questions or need clarification regarding this issue, they should contact Arrow's Customer Service Department at 1-800-523-8446.
- Quantity in commerce
- 188 units
- Distribution
- Worldwide Distribution -- USA, Belgium, Great Britain, Italy, Kazakhstan, Latvia, Russia, Slovenia, Turkey, and South Africa.
- Product codes and lots
- Lot Numbers: KF2073734 KS2083999 KF2083869 KF2083921
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1064-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1065-2013 | Class II | Terminated | Voluntary: Firm initiated |