Z-1569-2013Class II

Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of...

Terumo Cardiovascular Systems Corporation / initiated 2013-05-30 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of temperature-controlled water for blood heat exchangers... on . FDA classified it as Class II and gives the reason as "Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1569-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and is indicated for controlling and monitoring patient temperature
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Terumo sent beginning May 30, 2013, an URGENT MEDICAL DEVICE RECALL NOTIFICATION were sent to all customers with instructions to notify all customers of the recall requesting all affected units to be quarantined until the heatsink units are replaced Confirm receipt by filling out the response form containing communication by faxing or emailing the attached Affiliate Response Form to the fax number or email address indicated on the form. Questions should be directed to Terumo CVS customer service in the U.S.: Telephone: +1-734-663-4145; Fax: +1-734-741-6149
Quantity in commerce
7 units
Distribution
Worldwide Distribution including the countries of UAE (United Arab Emirates) HONG KONG and THAILAND. Not distributed in USA
Product codes and lots
Catalog Number 4416 and Serial Numbers: 4656, 4657, 4658,4659, 4661, 4662, and 4663.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1569-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1569-2013Class IITerminatedVoluntary: Firm initiated