Z-1782-2017Class II

Centricity PACS Web Diagnostic (WebDX) by GE Healthcare Dynamic Imaging Solutions is a device that receives...

GE Healthcare / initiated 2013-02-18 / Terminated / root cause: software

Ge Medical Systems Information T began a recall of Centricity PACS Web Diagnostic (WebDX) by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from... on . FDA classified it as Class II and gives the reason as "It was reported from a complaint that study category labels of "Prior" and "Baseline" and hanging position appeared incorrectly swapped when the hanging protocol for three comparison studies is manually applied from the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1782-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Centricity PACS Web Diagnostic (WebDX) by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from various imaging sources within the system or across computer networks at distributed locations.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
It was reported from a complaint that study category labels of "Prior" and "Baseline" and hanging position appeared incorrectly swapped when the hanging protocol for three comparison studies is manually applied from the patient folder.
Root cause tag
software
FDA root cause
Software design
Action
Firm sent notification letter to all impacted customer accounts. The letter provided the following instructions for customers to follow until the permanent fix for the issue was deployed: 1) The users should be instructed to look at the date on each exam displayed rather than just the labels assigned to the exam by the system. 2) If this problem is detected in your specific workflows, a system administrator can remove the study category labels (Current/Prior/Baseline)from all overlays on the images/exams. A software code change was deployed to customer sites to correct the issue.
Quantity in commerce
118
Distribution
Nationwide, Worldwide
Product codes and lots
Versions 1.0, 1.1, 2.0, 2.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1782-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1782-2017Class IITerminatedVoluntary: Firm initiated