Z-1473-2013Class II

SlingBar 350. Designed to meet the needs for lifting humans

Hill-Rom Inc. / initiated 2013-04-30 / Terminated / root cause: design

Hill-Rom, Inc. began a recall of SlingBar 350. Designed to meet the needs for lifting humans on . FDA classified it as Class II and gives the reason as "Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1473-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
SlingBar 350. Designed to meet the needs for lifting humans.
Product code
FSA Lift, Patient, Non-Ac-Powered
Reason for recall
Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.
Root cause tag
design
FDA root cause
Device Design
Action
On 4/30/13 an URGENT MEDICAL DEVICE RECALL CORRECTION was issued to all consignees detailing the hazards and actions that are needed. Included with this notification is a Response Form/Receipt and a Replacement Guide, complete and return the form to Hill-Rom as soon as possible. Hill-Rom will send you the applicable number of sling bars to you at no cost. After you receive the new sling bars, we request that you discard the old ones.
Quantity in commerce
3556
Distribution
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.
Product codes and lots
3156015, 3156002, 3156008' 3156004

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1473-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1473-2013Class IITerminatedVoluntary: Firm initiated
Z-1474-2013Class IITerminatedVoluntary: Firm initiated
Z-1475-2013Class IITerminatedVoluntary: Firm initiated
Z-1476-2013Class IITerminatedVoluntary: Firm initiated