SlingBar 360. Designed to meet the needs for lifting humans
Hill-Rom, Inc. began a recall of SlingBar 360. Designed to meet the needs for lifting humans on . FDA classified it as Class II and gives the reason as "Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1474-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- SlingBar 360. Designed to meet the needs for lifting humans.
- Product code
- FSA Lift, Patient, Non-Ac-Powered
- Reason for recall
- Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom is requesting products be removed from field use and replaced with an updated design.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 4/30/13 an URGENT MEDICAL DEVICE RECALL CORRECTION was issued to all consignees detailing the hazards and actions that are needed. Included with this notification is a Response Form/Receipt and a Replacement Guide, complete and return the form to Hill-Rom as soon as possible. Hill-Rom will send you the applicable number of sling bars to you at no cost. After you receive the new sling bars, we request that you discard the old ones.
- Quantity in commerce
- 16
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.
- Product codes and lots
- Product # 3156013
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1473-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1474-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1475-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1476-2013 | Class II | Terminated | Voluntary: Firm initiated |