Z-1114-2013Class II

CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to...

Roche Diagnostics GmbH / initiated 2013-03-06 / Terminated / root cause: design

Roche Diagnostics Operations, Inc. began a recall of CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT") on . FDA classified it as Class II and gives the reason as "Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1114-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
Product code
GJS Test, Time, Prothrombin
Reason for recall
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a day, seven days a week at 1-800-440-3638 for Accu check questions and 1-800-428-4674 for CoaguChek questions about the information contained in this UMDC.
Quantity in commerce
4467 both devices
Distribution
Nationwide Distribution.
Product codes and lots
catalog number 05530199160

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1114-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1113-2013Class IITerminatedVoluntary: Firm initiated
Z-1114-2013Class IITerminatedVoluntary: Firm initiated