Instrumentation Industries BE183-SU and BE 183 SU-12 Manual Jet Ventilators; BE 183-SUR and BE 183 SUR-12...
Instrumentation Industries Inc began a recall of Instrumentation Industries BE183-SU and BE 183 SU-12 Manual Jet Ventilators; BE 183-SUR and BE 183 SUR-12 Manual Jet Ventilators with Regulator and Gauge; BE... on . FDA classified it as Class II and gives the reason as "Firm initiated an update to Directions for Use necessitating replacement of previously distributed Directions for Use". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1125-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Instrumentation Industries, Incorporated
- Product
- Instrumentation Industries BE183-SU and BE 183 SU-12 Manual Jet Ventilators; BE 183-SUR and BE 183 SUR-12 Manual Jet Ventilators with Regulator and Gauge; BE 183-2 4 ft. Small Bore Tubing Assembly w/Inline Filter; and BE 183-R Regulator The BE 183-SUR and BE 183-SU Manual Jet Ventilators are designed to provide transtracheal ventilation in specific emergency situations of upper airway obstructions.
- Product code
- BTL Ventilator, Emergency, Powered (Resuscitator)
- Reason for recall
- Firm initiated an update to Directions for Use necessitating replacement of previously distributed Directions for Use.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Instrumentation Industries, Inc. initiated a recall of this product on September 7, 2012 by sending a letter via priority mail to all consignees. The letter identified the product, the problem, and the action to be taken by the customer. Consignees were instructed to replace the package insert of any affected product with the enclosed corrected package insert (now at Rev. H). The Directions for Use can alsbe downloaded from their website if they should need additional copies. If the product was further distributed consignees were asked to notify their consignees about this Field Correction. For questions regarding this recall call 412-854-1133.
- Quantity in commerce
- 2998
- Distribution
- Worldwide Distribution - USA including AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV. Shipped internationally to Canada, Ireland, Hong Kong, the Netherlands, New Zealand, and Thailand.
- Product codes and lots
- BE 183-2 -- lot numbers P0911307, P1109307, P1117907, P1210807, P1216107, P0112208, P0203508, P0208308, P0220008, P0313608, P0503608, P0510108, P0712408, P0816308, P0911708, P1008408, P0115009, P0203009, P0313109, P0603709, P0619009, P0806309, P0903509, P0905109, P0912309, P1005709, P1107809, P1205709, P1214309, P0111410, P0210810, P0212510, P0300210, P0314310, P0400710, P0404510, P0701210, P0705410, P0715010, P0816310, P0715010R10, P0705410R10, P1001810, P1011610, P1114810, P1204810, P1207310, P0109811, P0211311, P0404611, P0504911, P0603911, P0706511, P0801811, P0811811, P0818511, P0900611, P0908411, P0917111, P1008111, P1012211, P1113911, P1207111, P0107712, P0118212, P0211712, P0308512, P0409012, P0412912, P0802012, and P0813612; BE 183-R -- P0904007, P1015407, P1118107, P1209207, P0205608, P0208208, P0415408, P0509608, P0503008, P0517408R, P1215508, P0200709, P0308109, P0614909, P0712009, P0806409, P0903309, P1101909, P1101909R10, P0303710, P0409910, P0615710, P0710710, P0714510, P0814310, P0907010, P1214710, P0208711, P0208711R11, P1214710R11, P0716411, P0718111, P0807511, P0812211, P0816111, P0819511, P0914111, P1000511, P1015011, P1217911, P0117112, P0211312, P0313612, P0605312, P0709612, and P0807812; BE183-SU -- P0914607, P1015607, P1016907, P1119007, P1201807, P1211307, P0108808, P0112408, P0203008, P0208608, P0306908, P0313508, P0404908, P0510008, P0708808, P0806508, P0915908, P1020008, P1104508, P1109108, P1211608, P0102809, P0203109, P0208109, P0214609, P0302209, P0504909, P0603609, P0617209, P0618909, P0703509, P0804609, P0812609, P1005209, P1110009, P1221709, P0104510, P0108110, P0307110, P0411910, P0603510, P0616210, P0813510, P0912310, P1000910, P1017410, P0111111, P0207611, P0209411, P0216411, P0300611, P0309411, P0515711, P0619111, P0707411, P0717311, P0804011, P0809611, P0815311, P0901611, P0905011, P0915611, P1014511, P1103511, P1112811, P1116411, P1201911, P1213511, P1216511, P0103312, P0106212, P0114112, P0207712, P0211112, P0216312, P0320512, P0402512, P0408912, P0416912, P0509212, and P0513212; BE 183 SU-12 -- no lot numbers assigned; BE 183-SUR -- P0908607, P1006707, P1010807R, P1101207, P1117507, P1117707, P1210307, P1212007, P0103208, P0110308, P0116508, P0203208, P0214908, P0221508, P0317708, P0409808, P0409908, P0418308, P0508508, P0513308, P0601308, P0606608, P0704008, P0808108, P0606608R, P0814308, P0907308, P0916208, P1011108, P1016908, P1201708, P1208508, P0102409, P0113809, P0203709, P0102409R09, P0304509, P0405109, P0413009, P0501209, P0612709, P0709109, P0803109, P0910909, P1009609, P1106609, P1111509, P1206809, P1218009, P0114410, P0203010, P0304310, P0408410, P0500610, P0509410, P0607410, P0705710, P0711810, P0500610R10, P0805110, P0903110, P0909710, P1014710, P1102910, P0104411, P0116011, P0201211, P0206711, P0211011, P0219311, P0302911, P0308311, P0312211, P0400211, P0413811, P0506511, P0605911, P0609411, P0703111, P0708011, P0807911, P0809511, P0813811, P0819311, P0901311, P0905311, P0907511, P0912211, P1002711, P1018011, P1102911, P1104011, P1110711, P1207011, P0107812, P0115812, P0208312, P0210912, P0305212, P0309312, P0315812, P0400912, P0407212, P0417012, P0601712, P0701712, and P0712412; and BE 183 SUR-12 -- P1012107, P1102507, P1116107, P0319008, P1113408, P0201409, and P0703911.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1125-2013 | Class II | Terminated | Voluntary: Firm initiated |