Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers...

Zimmer GmbH / initiated 2013-02-08 / Terminated / root cause: packaging

Zimmer, Inc. began a recall of Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers: N1830R80-100, N1830R80-090, N1830R80-080, N1830R80-070... on . FDA classified it as Class II and gives the reason as "Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1045-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers: N1830R80-100, N1830R80-090, N1830R80-080, N1830R80-070, N1830R80-060, N1830R50-100, N1830R50-090, N1830R50-080, N1830R50-070, N1830R50-060, N1830L80-100, N1830L80-090, N1830L80-080, N1830L80-070, N1830L80-060, N1830L50-100, N1830L50-090, N1830L50-080, N1830L50-070, N1830L50-060, N182450-085, N182450-075, N182450-065, N182450-055, N182450-045. Retractor access system that allows placement of posterior spinal fixation implants.
Product code
HRX Arthroscope
Reason for recall
Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of product sterility) if proper aseptic technique is not followed when transferring the inner package into the sterile field. Patient infection may o
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Consignees were sent on 2/8/13 a Zimmer Spine "Urgent Medical Device Recall" letter dated February 08, 2013. The letter was sent to Surgeons and Facilities using the NexPosure MIS Access System. The letter described the product involved in the recall and the problem. Advised consignees to discontinue the use of the product and to contact their sales representative for removal. A consignee letter dated February 8, 2013 was also sent to Distributors, Sales Representatives, and Distribution Operation Managers Distributing the NexPosure MIS Access System. The letter contained instructions if they had the affected product or not and the steps to follow.
Quantity in commerce
58
Distribution
Nationwide Distribution including the states of AZ, CO, DE, FL, ID, LMA, MI, NH, NJ, NM, PA, RI, TX, and VA.
Product codes and lots
Lot Number 61225572  61664689  61664689  61657007  61657015  61657015  61267745  61267745  61267750  61267750  61267753  61267753  61657016  61657016  61801814  61895685  61657017  61657018  61657018  61267756  61267756  61267759  61267759  61267760  61267760  61267761  61267761  61267761  61873348  61657019  61657019  61657031  61657031

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1045-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1045-2013Class IITerminatedVoluntary: Firm initiated