Siemens Dimension(R) EXL(TM) LM (Siemens Material Number 10486890). Intended to measure a variety of analytes...
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension(R) EXL(TM) LM (Siemens Material Number 10486890). Intended to measure a variety of analytes in human body fluids on . FDA classified it as Class III and gives the reason as "Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1046-2013
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Dimension(R) EXL(TM) LM (Siemens Material Number 10486890). Intended to measure a variety of analytes in human body fluids.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
- FDA root cause
- Under Investigation by firm
- Action
- Siemens initiated their recall of this product on February 27, 2013 by sending an Important Customer Notification letter to consignees by FedEx. The letter identified the product, problem, and actions to be taken by the customers.
- Quantity in commerce
- 2207 total
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Siemens Material Number 10486890
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1046-2013 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1047-2013 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1048-2013 | Class III | Terminated | Voluntary: Firm initiated |