Z-0949-2013Class II

TotalCare SpO2RT2 bed

Hill-Rom Inc. / initiated 2013-02-26 / Terminated / root cause: software

Hill-Rom, Inc. began a recall of TotalCare SpO2RT2 bed. Product Usage: Bed, flotation therapy, powered bed may be used in a variety of settings including, but not limited to, acute care... on . FDA classified it as Class II and gives the reason as "In certain situations a software problem with the TotalCare SpO2RT 2 bed may occur that can lead to inadvertent loss of sufficient surface cushion pressure to adequately support the patient. This may increase the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0949-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
TotalCare SpO2RT2 bed. Product Usage: Bed, flotation therapy, powered bed may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare SpO2RT 2 Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.
Product code
IOQ Bed, Flotation Therapy, Powered
Reason for recall
In certain situations a software problem with the TotalCare SpO2RT 2 bed may occur that can lead to inadvertent loss of sufficient surface cushion pressure to adequately support the patient. This may increase the likelihood of skin irritation for patients at higher risk of developing pressure-related skin conditions. A correction to the software program is currently being developed and tested,
Root cause tag
software
FDA root cause
Software design
Action
Hill-Rom sent an Urgent Field Safey Notice letter dated February 26, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter advised customers to follow steps provided until the updated software is available. Customers were requested to forward a copy of this letter to any other facility personnel you deem appropriate. Questions should be directed to Hill-Rom Technical Support at 800-445-3720
Quantity in commerce
309
Distribution
USA Nationwide Distribution including the state of Canada
Product codes and lots
S/N between N230AM1358 & O023AM3687.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0949-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0949-2013Class IITerminatedVoluntary: Firm initiated