Z-1122-2013Class II
ADVIA Centaur Vitamin D Calibrator 6-pack; Reference Number:10630911. For use in the quantitative...
Siemens Healthcare Diagnostics Inc. / initiated 2013-03-13 / Terminated
Siemens Healthcare Diagnostics, Inc began a recall of ADVIA Centaur Vitamin D Calibrator 6-pack; Reference Number:10630911. For use in the quantitative determination of Vit D on . FDA classified it as Class II and gives the reason as "Negative bias in patient samples for Vitamin D". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1122-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- ADVIA Centaur Vitamin D Calibrator 6-pack; Reference Number:10630911. For use in the quantitative determination of Vit D.
- Product code
- JIS Calibrator, Primary
- Reason for recall
- Negative bias in patient samples for Vitamin D
- FDA root cause
- No Marketing Application
- Action
- Siemens Healthcare issued Urgent Medical Device Recall was sent to all affected Siemens Healthcare Diagnostics customers in the United States on March 13th, 2013. This notice informs the customer of the recall and the actions that need to be taken. The Urgent Medical Device Recall instructs customers to discontinue the use of ADVIA Centaur VitD Assay Kit Lots (REF 10491994 and 10631021) and ADVIA Centaur Systems VitD Cal (REF 10493589 and 10630911) released prior to the recalibration of the assay and request replacement lots. This notice also directs customers to review the contents of the recall notice with their laboratory or medical director. A confirmation fax-back form is included to ensure that all customers in the United States have been notified and have complied with the mandatory action.
- Quantity in commerce
- 107 6pk Cals
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot Number: C3410 Kit Lot Number/Exp. Date: 95010A10 8/16/2013
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1119-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1120-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1121-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1122-2013 | Class II | Terminated | Voluntary: Firm initiated |