Stryker Orthopaedics Baseplate Impactor Extractor TRIATHLON Instruments. Howmedica Osteonics Corp. 325...
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Baseplate Impactor Extractor TRIATHLON Instruments. Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah, NJ 07430 USA A subsidiary of... on . FDA classified it as Class II and gives the reason as "Stryker received reports from the field indicating that the handle detached from the cam of the Baseplate Impactor/Extractor Assembly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1061-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics Baseplate Impactor Extractor TRIATHLON Instruments. Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah, NJ 07430 USA A subsidiary of Stryker Corporation. Made in USA. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France The Baseplate Impactor Extractor is intended to securely attach to the tibial baseplate or baseplate trial for impaction and extraction during surgery.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Stryker received reports from the field indicating that the handle detached from the cam of the Baseplate Impactor/Extractor Assembly.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker sent Notification Letters/Product Accountablility Forms dated March 1, 2013 via FedEx to their Branches/Agencies on March 1, 2013.. Notification Letters/Product Recall Acknowledgement Forms dated March 4, 2013, were sent via FedEx to Hospital Risk Management on March 4, 2013. The Notification letters identified the product, the problem, and instructed customers on how to identify affected product; informs customers of the issues and potential hazards with the affected product and instructs them to contact their representative to arrange for return/replacement of any affected product. Customers are also asked to fax back the provided "Urgent Medical Device Removal Notification Acknowledgement Form" to Stryker within 5 days of receipt. Customers with questions were instructed to call 201-972-2100. For questions regarding this recall call 201-831-5523.
- Quantity in commerce
- 871 units
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Japan, Australia, China, New Zealand, France, Italy, Netherlands, Poland, Spain, Sweden, Switzerland and the United Kingdom.
- Product codes and lots
- Catalog Number 6541-4-805. Lot Numbers: RD4V204; RD5V234; RD5V235; RD5V235A; RD5V235Y; RD5V236; RD5V236A; RD5V237; RD6A009; RD6A010; RD6A011; RD6A011P; RD6A012; RD6A051; RD6A052; RD6A053; RD6A053M; RD6A053T; RD6A061; RD6A061A; RD6A062; RD6A062N; RD6A063; RD6A063T; RD6A065; RD6A065D; RD6A065T; RD6A066; RD6A066A; RD6C103; RD6C103L; RD6C104; RD6C104Y; RD6C105; RD6C105M; RD6C105T; RD6C106; RD6C106A; RD6C108; RD6C109; RD6C109L; RD6C126; RD6H178; RD6H184; RD6H187; RD6L222; RD6L223; RD6L223A; RD6L224; RD6L224A; RD6L226; RD6L227; RD6L227A; RD6L227N; RD6M148; RD6M158; RD6M159; RD6M161; RD6M163; RD6M163A; RD6M163L; RD6M164; RD6M164A; RD6M165; RD6N046; RD6N046L; RD6N047; RD6N047N; RD6N048; RD6N052; RD6N052T; RD6N053; RD6N054; RD6N072; RD6N072H; RD6N073; RD6N074; RD6N074J; RD6N158; RD6N159; RD6N160; RD6S054; RD6S055; RD6S055A; RD6S056; RD6S081; RD6S082; RD6S082Y; RD6S083; RD6S083Y; RD6S084; RD6S084A; RD6S085; RD6S085Y; RD6S142; RD6S142Y; RD6S143; RD6S143A; RD6S144; RD6S144Y.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1061-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.