Z-1061-2013Class II

Stryker Orthopaedics Baseplate Impactor Extractor TRIATHLON Instruments. Howmedica Osteonics Corp. 325...

Howmedica Osteonics Corp / initiated 2013-02-23 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Baseplate Impactor Extractor TRIATHLON Instruments. Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah, NJ 07430 USA A subsidiary of... on . FDA classified it as Class II and gives the reason as "Stryker received reports from the field indicating that the handle detached from the cam of the Baseplate Impactor/Extractor Assembly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1061-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics Baseplate Impactor Extractor TRIATHLON Instruments. Howmedica Osteonics Corp. 325 Corporate Drive, Mahwah, NJ 07430 USA A subsidiary of Stryker Corporation. Made in USA. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 MEYZIEU Cedex France The Baseplate Impactor Extractor is intended to securely attach to the tibial baseplate or baseplate trial for impaction and extraction during surgery.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Stryker received reports from the field indicating that the handle detached from the cam of the Baseplate Impactor/Extractor Assembly.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stryker sent Notification Letters/Product Accountablility Forms dated March 1, 2013 via FedEx to their Branches/Agencies on March 1, 2013.. Notification Letters/Product Recall Acknowledgement Forms dated March 4, 2013, were sent via FedEx to Hospital Risk Management on March 4, 2013. The Notification letters identified the product, the problem, and instructed customers on how to identify affected product; informs customers of the issues and potential hazards with the affected product and instructs them to contact their representative to arrange for return/replacement of any affected product. Customers are also asked to fax back the provided "Urgent Medical Device Removal Notification Acknowledgement Form" to Stryker within 5 days of receipt. Customers with questions were instructed to call 201-972-2100. For questions regarding this recall call 201-831-5523.
Quantity in commerce
871 units
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Canada, Japan, Australia, China, New Zealand, France, Italy, Netherlands, Poland, Spain, Sweden, Switzerland and the United Kingdom.
Product codes and lots
Catalog Number 6541-4-805. Lot Numbers: RD4V204; RD5V234; RD5V235; RD5V235A; RD5V235Y; RD5V236; RD5V236A; RD5V237; RD6A009; RD6A010; RD6A011; RD6A011P; RD6A012; RD6A051; RD6A052; RD6A053; RD6A053M; RD6A053T; RD6A061; RD6A061A; RD6A062; RD6A062N; RD6A063; RD6A063T; RD6A065; RD6A065D; RD6A065T; RD6A066; RD6A066A; RD6C103; RD6C103L; RD6C104; RD6C104Y; RD6C105; RD6C105M; RD6C105T; RD6C106; RD6C106A; RD6C108; RD6C109; RD6C109L; RD6C126; RD6H178; RD6H184; RD6H187; RD6L222; RD6L223; RD6L223A; RD6L224; RD6L224A; RD6L226; RD6L227; RD6L227A; RD6L227N; RD6M148; RD6M158; RD6M159; RD6M161; RD6M163; RD6M163A; RD6M163L; RD6M164; RD6M164A; RD6M165; RD6N046; RD6N046L; RD6N047; RD6N047N; RD6N048; RD6N052; RD6N052T; RD6N053; RD6N054; RD6N072; RD6N072H; RD6N073; RD6N074; RD6N074J; RD6N158; RD6N159; RD6N160; RD6S054; RD6S055; RD6S055A; RD6S056; RD6S081; RD6S082; RD6S082Y; RD6S083; RD6S083Y; RD6S084; RD6S084A; RD6S085; RD6S085Y; RD6S142; RD6S142Y; RD6S143; RD6S143A; RD6S144; RD6S144Y.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1061-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1061-2013Class IITerminatedVoluntary: Firm initiated