Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray...
Siemens Medical Solutions USA, Inc began a recall of Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system on . FDA classified it as Class II and gives the reason as "The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0997-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent a Customer Safety Advisory Notice on March 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to pass the information on to all those who need to be aware of it and to instruct their personnel accordingly. Customers were asked to forward the safety information to other organizations that couald be affected by this action. For questions regarding this recall call 610-219-6300.
- Quantity in commerce
- 81
- Distribution
- Nationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.
- Product codes and lots
- Model numbers 10094135, 10094137, 10094139, 10094141 and 10280959.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0997-2013 | Class II | Terminated | Voluntary: Firm initiated |