TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc...
Smith & Nephew Inc began a recall of TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery on . FDA classified it as Class II and gives the reason as "One batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1132-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- One batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter).
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Smith & Nephew sent an Urgent Product Recall Notification letter dated March 6, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to cease use and distribution of the affected product and quarantine for return. Non-responding consignees were notified again on March 15, 2013, by email. Surgeons were to be notified by letter beginning on March 20, 2013. For questions regarding this recall call 978-749-1330.
- Quantity in commerce
- 23 units
- Distribution
- Nationwide Distribution including CA, FL, GA, MA, MI, NC, SC, OH, and MO.
- Product codes and lots
- Lot Number 12FM12182
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1132-2013 | Class II | Terminated | Voluntary: Firm initiated |