Z-1124-2013Class II

REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty...

Biomet U.K., Ltd. / initiated 2013-03-20 / Terminated / root cause: design

Biomet U.K., Ltd. began a recall of REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical... on . FDA classified it as Class II and gives the reason as "Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1124-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Biomet U.K., Ltd.
Product
REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.
Product code
NRA Prosthesis, Knee, Femorotibial, Unicompartmental, Semi-Constrained, Metal/Polymer, Mobile Bearing
Reason for recall
Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
Root cause tag
design
FDA root cause
Device Design
Action
BIOMET sent an Urgent Medical Device Recall Notice on March 20, 2013, to all affected customers with directions to forward the notice if product was further distributed to Hospital Operating Room managers. The notice identified the recalled device, reason for the recall, and possible adverse events directing consignees to locate and discontinue use of the device. A FAX back Response form was to be returned to 574-372-1683 within 3 days of receipt prior to return of the product to Biomet, USA. Questionsregarding this recall should be addressed to 574 372-1570 M-F, 8a.m. -5 p.m.
Quantity in commerce
430
Distribution
Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
Product codes and lots
32-422760 ( sold as part of instrument set 32-422763 ) ZB100901,ZB110201,ZB110301,ZB110302,ZB110303, ZB110304,ZB110601,ZB110602,ZB110701,ZB111201, ZB111202,ZB120201,ZB120202,ZB120701,ZB120801, ZB120802,ZB121102,ZB121103

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1124-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1124-2013Class IITerminatedVoluntary: Firm initiated