Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in...
Stryker Endoscopy began a recall of Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during... on . FDA classified it as Class II and gives the reason as "The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123C , instead of the correct 132-133C". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1224-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
- Product code
- GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered
- Reason for recall
- The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123C , instead of the correct 132-133C.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Stryker sent "Urgent Device Correction" letters dated March 27, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
- Quantity in commerce
- 147,921 - all sizes
- Distribution
- Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
- Product codes and lots
- Manufacturer Part Number: 250-070-411
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1218-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1219-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1220-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1221-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1222-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1223-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1224-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1225-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1226-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1227-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1228-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1229-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1230-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1231-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1232-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1233-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1234-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1235-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1236-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1237-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.