Z-1152-2013Class II

VITROS Chemistry Products PHBR (Phenobarbital) Slides, REF 822 1384. For the quantitative measure of...

Ortho-Clinical Diagnostics, Inc. / initiated 2013-03-14 / Terminated

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products PHBR (Phenobarbital) Slides, REF 822 1384. For the quantitative measure of phenobarbital (PHBR) concentration in serum and plasma on . FDA classified it as Class II and gives the reason as "Ortho Clinical Diagnostics (OCD) received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using VITROS Chemistry Products PHBR...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1152-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products PHBR (Phenobarbital) Slides, REF 822 1384. For the quantitative measure of phenobarbital (PHBR) concentration in serum and plasma.
Product code
DLZ Enzyme Immunoassay, Phenobarbital
Reason for recall
Ortho Clinical Diagnostics (OCD) received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using VITROS Chemistry Products PHBR Slides. OCD's investigation confirmed that positively biased values for patient samples could be obtained that may not be detected by quality control fluids.
FDA root cause
Unknown/Undetermined by firm
Action
On 3/14/13, URGENT PRODUCT CORRECTION NOTIFICATION letters (dated 3/14/13), Questions & Answers, and Response Forms were sent to consignees via FedEx courier or US Postal Service Priority Mail. Consignees were informed of the issue and advised to immediately discontinue using VITROS PHBR Slides. On 3/14/13, distributor letters were sent via US Postal Service Priority Mail to the distributors. They were advised of the issue and instructed to discontinue distributing and discard their remaining inventory of VITROS PHBR Slides. On 3/14/13, foreign affiliates were informed by e-mail of the issue and instructed to notify their consignees of the issue and actions.
Quantity in commerce
Domestic: 11,164 units, Foreign: 594 units
Distribution
Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
Product codes and lots
All Lot Numbers containing: GEN 38, Coating No. 0060, Expiration Date Range 03/01/13 through 07/01/13; GEN 39, Coating No. 0062, Expiration Date Range 07/01/13 through 09/01/13; GEN 40, Coating No. 0063, Expiration Date Range 10/01/13 through 12/01/13; GEN 41, Coating No. 0064, Expiration Date Range 12/01/13 through 02/01/14. [A Coating (CTG) is a unique identifier within the Lot Number of some of OCD's products. The term GEN and CTG refer to a specific segments of the reagent lot number. Current lots within expiry were all from Coatings 0060, 0062, 0063 or 0064.].

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1152-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1152-2013Class IITerminatedVoluntary: Firm initiated