Z-1149-2013Class II

MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0)

Maquet Cardiovascular LLC / initiated 2013-03-07 / Terminated / root cause: software

Maquet Cardiovascular, LLC began a recall of MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is a blood oxygenation and carbon dioxide... on . FDA classified it as Class II and gives the reason as "It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1149-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular LLC
Product
MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours).
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl
Root cause tag
software
FDA root cause
Software change control
Action
MAQUET sent an Urgent Medical Device Field Correction letters dated March 7, 2013 via Federal Express to all affected customers. The letter identifies the affected product, problem and actions to be taken. Customers are asked to complete and submit the included Customer Fax-back Form. Maquet Service will exchange all affected products with the updated version. For questions MAQUET Customer Service at 888-627-8383 (press option 2, followed by option 2).
Quantity in commerce
154 units affected
Distribution
Worldwide Distribution - USA Nationwide and the following countries: Australia, Austria, Belarus, Belgium, Canada, Columbia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Iceland, Iran, Ireland, Italy, Jordan, Kazakhstan, Korea, Kuwait, Luxembourg, The Netherlands, Norway, Oman, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and the United Kingdom.
Product codes and lots
Catalog Number: 70104.8012 Cardiohelp-i. Multiple serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1149-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1149-2013Class IITerminatedVoluntary: Firm initiated