Z-1167-2013Class II

AU5800 Clinical Chemistry Analyzer The Beckman Coulter AU5800 Clinical Chemistry Analyzer is an automated...

Siemens Healthcare Diagnostics Inc. / initiated 2013-03-20 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of AU5800 Clinical Chemistry Analyzer The Beckman Coulter AU5800 Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples... on . FDA classified it as Class II and gives the reason as "Beckman Coulter, Inc. initiated this recall because there is a software error that has the potential to cause the reporting of erroneous patient results because of misidentification by the AU5800 system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1167-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AU5800 Clinical Chemistry Analyzer The Beckman Coulter AU5800 Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. Applications include colorimetric, turbidimetric, latex agglutination, homogeneous enzyme immunoassay, and in selective electrode.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter, Inc. initiated this recall because there is a software error that has the potential to cause the reporting of erroneous patient results because of misidentification by the AU5800 system.
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter sent an Urgent Product Correction letter during the week of March 18, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were informed of the software limitation and provides instructions to be taken (temporary workaround) until the software is upgraded and released. Customer contact information provided: "Call Center Hotline at 800 854-3633. New customers will receive a copy of the customer notification at the time of installation by the Field Service Personnel until revised software that corrects this issue is released.
Quantity in commerce
490
Distribution
Worldwide Distribution - USA, including Canada and Puerto Rico and the states of NJ, TX, CA, UT, KS, MA, NY, MI, GA, VA, IN, WI, IL, OH, NV, NC and PA. and the countries of Australia, Belgium, Czech Republic, Japan, Germany, Brazil, China, Italy, Ireland, Israel, India, Hong Kong, Malaysia, Netherlands, Spain, Switzerland, France, Taiwan, Turkey, United Kingdom and Vietnam.
Product codes and lots
All Instrument Serial Numbers with Software Versions 4.11 and Prior  Model Numbers AU5811-01, AU5821-01, AU5831-01, AU5841-01, AU5811-02, AU5821-02, AU5831-02, AU5841-02, AU5811-03, AU5821-03, AU5831-03, AU5841-03, AU5811-04, AU5821-04, AU5831-04, AU5841-04, AU5811-06, AU5821-06, AU5831-06, AU5841-06  Part Numbers A94906, A94911, A94916, A91921 A94907, A94912, A94917, A94922 A94908, A94913, A94918, A94923 A94909, A94914, A91949, A94924 A94910, A94915, A94920, A94925  Model Numbers and Part Numbers listed above respecitively.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1167-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1167-2013Class IITerminatedVoluntary: Firm initiated