Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1055-2013Class II

Glow 'N Tell 20 cm Tape: 1100-00 (100 strips); 1100-50 (50 strips); 1100-20 (20 strips). Intended to be...

LeMaitre Vascular Inc. / initiated 2013-03-18 / Terminated / root cause: packaging

LeMaitre Vascular, Inc. began a recall of Glow 'N Tell 20 cm Tape: 1100-00 (100 strips); 1100-50 (50 strips); 1100-20 (20 strips). Intended to be placed on the skin to assist during imaging procedures on . FDA classified it as Class II and gives the reason as "Devices were not sealed correctly during the manufacturing process, and the sterility of these products have been compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1055-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LeMaitre Vascular Inc.
Product
Glow 'N Tell 20 cm Tape: 1100-00 (100 strips); 1100-50 (50 strips); 1100-20 (20 strips). Intended to be placed on the skin to assist during imaging procedures.
Product code
KFX Assembly, Thigh/Knee/Shank/Ankle/Foot, External
Reason for recall
Devices were not sealed correctly during the manufacturing process, and the sterility of these products have been compromised.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
LeMaitre sent an "URGENT:VASCUTAPE RADIOPAQUE TAPE DEVICE FIELD SAFETY NOTICE" dated March 20, 2013 to affected customers. The recall letter contains a form that requests customers to return to LeMaitre Vascular, Inc. as a record of notification and reconciliation. The letter provides instructions on how the customer can inspect the defective tape seals and/or return the affected products for replacement/credit.
Quantity in commerce
13560
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, EU, and Asia.
Product codes and lots
SGL1190, exp 2017-02; SGL1196, exp 2017-09; SGL1197, exp. 2017-09

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1055-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1054-2013Class IITerminatedVoluntary: Firm initiated
Z-1055-2013Class IITerminatedVoluntary: Firm initiated