Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system...
Siemens Medical Solutions USA, Inc began a recall of Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(R) Workflow SLR is a digital radiology... on . FDA classified it as Class II and gives the reason as "Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1105-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(R) Workflow SLR is a digital radiology information system (RIS) with integrated modules for patient administration, examination, reporting, statistics and system administration, as well as multiple interfaces for patient registration, order entry, and external billing modules. It is not intended to treat disease or administer any medicinal substances or drugs. There is no contact between device and patient. syngo(R) Workflow SLR can be used as a separate RIS, interfacing to a Hospital Information System (HIS) or Picture Archiving and Communication Systems (PACS
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or auto displayed, dismissed by the user, and then a different patient and signed exam is selected in single patient view. The Notes Viewer will
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical sent a Customer Safety Advisory Notice letter dated October 5, 2012 , to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. There will be a Hot Fix available. If you have any further questions please contact Siemens Customer Product Support at (1-888) 767-8326.
- Quantity in commerce
- 37
- Distribution
- Worldwide Distribution - USA including the states of AL, CA, CO, DC, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WI, WV, and WY. and the countries of Canada and Australia.
- Product codes and lots
- Model numbers 10558922, 10558933 and 10558980 with software version VA31A SP1/SP2/SP3
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1105-2013 | Class II | Terminated | Voluntary: Firm initiated |