Z-1105-2013Class II

Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system...

Siemens Medical Solutions USA, Inc. / initiated 2012-10-05 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(R) Workflow SLR is a digital radiology... on . FDA classified it as Class II and gives the reason as "Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1105-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(R) Workflow SLR is a digital radiology information system (RIS) with integrated modules for patient administration, examination, reporting, statistics and system administration, as well as multiple interfaces for patient registration, order entry, and external billing modules. It is not intended to treat disease or administer any medicinal substances or drugs. There is no contact between device and patient. syngo(R) Workflow SLR can be used as a separate RIS, interfacing to a Hospital Information System (HIS) or Picture Archiving and Communication Systems (PACS
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or auto displayed, dismissed by the user, and then a different patient and signed exam is selected in single patient view. The Notes Viewer will
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical sent a Customer Safety Advisory Notice letter dated October 5, 2012 , to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. There will be a Hot Fix available. If you have any further questions please contact Siemens Customer Product Support at (1-888) 767-8326.
Quantity in commerce
37
Distribution
Worldwide Distribution - USA including the states of AL, CA, CO, DC, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WI, WV, and WY. and the countries of Canada and Australia.
Product codes and lots
Model numbers 10558922, 10558933 and 10558980 with software version VA31A SP1/SP2/SP3

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1105-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1105-2013Class IITerminatedVoluntary: Firm initiated