Z-0773-2014Class II

Imagecast PACS is an image management system intended to be used by trained professionals, including but not...

GE Healthcare / initiated 2011-03-04 / Terminated / root cause: software

Ge Healthcare It began a recall of Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical... on . FDA classified it as Class II and gives the reason as "In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0773-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image and interpretations. Mammongraphic may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas
Root cause tag
software
FDA root cause
Software design
Action
Important Product Information Letters dated March 4, 2011 were sent to all customers that received the defective software. The letters advised customers to upgrade their PACS software version to Imagecast 3.6. The software upgrade can be requested from GE Support via the eService portal or by telephone at 1-800-437-1171. If the customer elects to not upgrade the software, instructions were provided for a new system calibration procedure.
Quantity in commerce
43 units
Distribution
Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.
Product codes and lots
Imagecast PACS versions prior to 3.6.51.01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0773-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0773-2014Class IITerminatedVoluntary: Firm initiated