Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1238-2013Class II

Synthes Radial Head Prosthesis: Trial Radial Head Elbow joint prosthesis: Synthes Radial Head Prosthesis...

Synthes GmbH / initiated 2013-02-22 / Terminated / root cause: design

Synthes USA HQ, Inc. began a recall of Synthes Radial Head Prosthesis: Trial Radial Head Elbow joint prosthesis: Synthes Radial Head Prosthesis System is a two-piece modular system comprised of... on . FDA classified it as Class II and gives the reason as "The Trial Head may come loose from the implant stem during manipulation of the arm during surgery. To date, no adverse events have been reported related to this issue and this recall is not being initiated as a result...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1238-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Radial Head Prosthesis: Trial Radial Head Elbow joint prosthesis: Synthes Radial Head Prosthesis System is a two-piece modular system comprised of titanium alloy stem and cobalt chrome components with an integral screw and side-loading application to allow in-situ assembly.
Product code
KWI Prosthesis, Elbow, Hemi-, Radial, Polymer
Reason for recall
The Trial Head may come loose from the implant stem during manipulation of the arm during surgery. To date, no adverse events have been reported related to this issue and this recall is not being initiated as a result of adverse events.
Root cause tag
design
FDA root cause
Device Design
Action
Synthes sent an Urgent Notice: Medical Device Recall letter dated February 22, 2013, to all affected consignees. The letter requested consignees examine their inventory, remove them from use and return them to Synthes. Customers were asked to complete the attached Verification Section at the end of the letter indicating whether or not they have any of the affected product. For questions customers were instructed to call 610-719-5450 or email [email protected]. For questions regarding this recall call 610-719-5000.
Quantity in commerce
417
Distribution
Nationwide Distribution: USA including states of: IL, IN, MA, MI, MO, NJ, NY, OH, OR, PA, SC, TX, VA, WA and WI.
Product codes and lots
All Lots of Part Numbers: 03.402.018, 03.402.020, 03.402.022, 03.402.024, 03.402.026, 03.402.028, 03.402.218, 03.402.220, 03.402.222, 03.402.224, 03.402.226, 03.402.228, 03.402.418, 03.402.420, 03.402.422, 03.402.424, 03.402.426, 03.402.428, 03.402.618, 03.402.620, 03.402.622, 03.402.624, 03.402.626, and 03.402.628.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1238-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1238-2013Class IITerminatedVoluntary: Firm initiated