The Centricity Laboratory System is intended to be an information system designed to support the clinical and...

GE Healthcare / initiated 2012-11-20 / Terminated / root cause: software

GE Healthcare It began a recall of The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the... on . FDA classified it as Class II and gives the reason as "If environment variables MICRO_CORRECT_MSG_TYPE and CORRECT_MSG_TYPE are set to different values, in Microbiology result entry when released results (Final Report released) are changed and the Corrected Result messages...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1585-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network. This software system will not be the software directly used to run, steer, or control any specific laboratory analyzer or equipment. It is intended for use in laboratory facilities, including a central laboratory or in a multiple laboratory environment servicing satellite laboratories, reference laboratories, clinics, etc. Centricity Laboratory System is not intended for direct patient contact.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
If environment variables MICRO_CORRECT_MSG_TYPE and CORRECT_MSG_TYPE are set to different values, in Microbiology result entry when released results (Final Report released) are changed and the Corrected Result messages are edited on the free-text form, some of the originally entered panel free-text is lost. For this issue to occur the original result must be amended and the environmental variabl
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated November 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Software correction is available. Please contact your GE service representative to request or install this update. Please contact your GE Service Representative with any question regarding this issue. For other questions please contact GE Helpdesk. Help Desk: North America: 1-888-778-3375 Help Desk: Europe: + 44 1603 877595 Help Desk: Asia Pacific + 61 3 5327 4000
Quantity in commerce
9 Units
Distribution
Worldwide Distribution - US Distribution including the states of OH, NY, GA, TX, TN and MN and the countries of Australia, Qatar and England
Product codes and lots
Software Versions 3.3, 4.0, and 4.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1585-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1585-2013Class IITerminatedVoluntary: Firm initiated