Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an information...
GE Healthcare It began a recall of Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an information system designed to support the clinical and... on . FDA classified it as Class II and gives the reason as "The Centricity Laboratory removes free-text notes on a master panel when the user deletes the slave panel. The deletion of a slave panel will (auto slash) the corresponding test (item) results on the master panel...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2671-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- The Centricity Laboratory removes free-text notes on a master panel when the user deletes the slave panel. The deletion of a slave panel will (auto slash) the corresponding test (item) results on the master panel. Additionally, the displayed and HL7 results show neither the appropriate abnormal flag value nor the reference range values, when using large_num.g >7 to report numeric results.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated Nov 20, 2012 to all affected customers. The letter described the product, problems and actions to be taken. The customers were instructed to follow the safety instructions and to ensure that all potential users in their facility are made aware of this safety notification and the recommended actions. The software has been modified to correct these defects. The modified software will be released and installed in user sites through GE Healthcare FMI #86071. Customers were instructed to contact their GE Service Representative with any questions regarding this issue. For other questions, contact GE Helpdesk: North America; (905) 479-1249 or 1-888-778-3375; email: [email protected] Europe; +44 1603 877595; email: [email protected] Asia Pacific; +61 3 5327 4000; email: [email protected]. For questions regarding this recall call 847-277-5000.
- Quantity in commerce
- 17
- Distribution
- Worldwide Distribution - US Distribution to GA, OH, NY, TX, TN ,MN and Internationally to India, Australia, Canada, England, Qatar, and Scotland.
- Product codes and lots
- Versions 4.1, 4.0, 3.3.2 & 3.3Almp
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2671-2016 | Class II | Terminated | Voluntary: Firm initiated |