Z-1164-2013Class II

GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and...

GE Healthcare / initiated 2011-10-06 / Terminated / root cause: software

GE Healthcare, LLC began a recall of GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras... on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1164-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration
Product code
LLZ System, Image Processing, Radiological
Reason for recall
GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts in attenuation corrected images. This may lead to a potential for misdiagnosis.
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated October 5, 2011, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees were instructed to contact their local GE Healthcare Service Representative if they have questions or concerns. They may also contact GE Healthcare Call Center , United States or Canada at 800-437-1171. Other countries 262-896-2890. For questions regarding this recall call 262-513-4122.
Quantity in commerce
11
Distribution
Worldwide Distribution - USA including GA, and NY; and countries of: Canada, Israel, Italy and Switzerland.
Product codes and lots
Mfg Lot or Serial # 0000CZC7105FJM000 00000CZC827NRN000 00000CZC8283V4000 0000CZC8283V47000 0000CZC8283V72001 0000CZC8283TX5001 00000000211423001 0000CZC8161XJT000 0000CZC94154WV003 0000CZC83942HL000 0000CZC8396CC9001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1164-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1164-2013Class IITerminatedVoluntary: Firm initiated