GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and...
GE Healthcare, LLC began a recall of GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras... on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1164-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts in attenuation corrected images. This may lead to a potential for misdiagnosis.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated October 5, 2011, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees were instructed to contact their local GE Healthcare Service Representative if they have questions or concerns. They may also contact GE Healthcare Call Center , United States or Canada at 800-437-1171. Other countries 262-896-2890. For questions regarding this recall call 262-513-4122.
- Quantity in commerce
- 11
- Distribution
- Worldwide Distribution - USA including GA, and NY; and countries of: Canada, Israel, Italy and Switzerland.
- Product codes and lots
- Mfg Lot or Serial # 0000CZC7105FJM000 00000CZC827NRN000 00000CZC8283V4000 0000CZC8283V47000 0000CZC8283V72001 0000CZC8283TX5001 00000000211423001 0000CZC8161XJT000 0000CZC94154WV003 0000CZC83942HL000 0000CZC8396CC9001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1164-2013 | Class II | Terminated | Voluntary: Firm initiated |