Z-0337-2014Class II

Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare...

GE Healthcare / initiated 2013-03-11 / Terminated / root cause: software

GE Healthcare It began a recall of Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery systems such as a Labor and Delivery... on . FDA classified it as Class II and gives the reason as "1) Under rare network conditions, a Centricity Perinatal process may maintain an incorrect count of Patient slots in use which may lead to an inability to access the patient roster. The user is temporarily unable to...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0337-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery systems such as a Labor and Delivery, Postpartum Maternal Care, newborn Nursery, critical care units including Neonatal Intensive Care Units (NICU), and may also be used in physicians offices and outpatient clinics. This product is primarily intended to serve the purpose of electronic documentation of clinical data and is designed to accept, transfer, display, calculate, store and manage clinical data. System capabilities provide the user with the ability to acquire data from the medical devices and to document, annotate, display, store, print, export and retrieve patient clinical practice information. This product is intended for professional use only.
Product code
HGM System, Monitoring, Perinatal
Reason for recall
1) Under rare network conditions, a Centricity Perinatal process may maintain an incorrect count of Patient slots in use which may lead to an inability to access the patient roster. The user is temporarily unable to review or edit form and chart information. This could result in a potential delay of care; however active fetal strips and permanent patient data storage are not affected. 2) When
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare sent an URGENT MEDICAL DEVICE CORRECTION letter dated March 11, 2013, to all affected customers. The letters requested that direct accounts ensure that all potential users of the software were made aware of the potential problems and the recommended actions. Direct accounts were asked to return a customer response form and to return it to GE Healthcare via fax at 847-939-1522 or e-mail at [email protected]. Version 6.94 of the software has been modified to correct these defects. GE Technical Support: 800-433-2009.
Quantity in commerce
1,132 units
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Australia, Canada, Denmark, Germany, Hong Kong, Korea, South Korea, Thailand and United Kingdom.
Product codes and lots
Centricity Perinatal and Centricity Intensive Care Versions: 6.60, 6.60.9, 6.70.0, 6.70.1, 6.70.5, 6.70.6, 6.80.0, 6.80.1, 6.90.0, 6.90.1, 6.91.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0337-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0337-2014Class IITerminatedVoluntary: Firm initiated