Z-1187-2013Class II

Rad-8 Pulse Oximeters The Rad-8 family of pulse oximeters and accessories are indicated for the continuous...

Masimo Corporation / initiated 2013-03-13 / Terminated / root cause: design

Masimo Corporation began a recall of Rad-8 Pulse Oximeters The Rad-8 family of pulse oximeters and accessories are indicated for the continuous, noninvasive monitoring of functional oxygen... on . FDA classified it as Class II and gives the reason as "Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off without the operator pressing the power button". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1187-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Masimo Corporation
Product
Rad-8 Pulse Oximeters The Rad-8 family of pulse oximeters and accessories are indicated for the continuous, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor).
Product code
DQA Oximeter
Reason for recall
Masimo has identified a very small number (0.02%) of Rad-8 devices with 20-pin patient cable connector that, when moved, can power off without the operator pressing the power button.
Root cause tag
design
FDA root cause
Device Design
Action
Masimo Corporation sent an Urgent Recall Notification letter dated March 8, 2013, to all affected customers The letter identified the product the problem and the action needed to be taken by the customer. 1.Customers were instructed to please promptly remove all Rad-8 device(s) from use and your inventory and segregate these devices. 2. Please call 1-800-326-4890 and select option 2 for Technical Services. Determine with Technical Services whether you will return the affected device(s) to Masimo for repair or if eligible, receive a repair kit. If you will return the device(s) for repair, obtain a Return Material Authorization Number (RMA number). 3. Complete the Tracking/Verification Form in Attachment 2 and fax it to Masimo at 1-949-297-7700, even if you do not have or intend to use the affected device(s). 4. If you are eligible and intend to repair units onsite fax the completed original form in Attachment 2 to Masimo at 1-949-297-7700. 5. If you have elected to return the affected device(s), and have an RMA number include the completed original Form in Attachment 2 with the returned device(s) and indicate the RMA number on the outside of the return-shipping box. We apologize for the inconvenience. Please be assured that Masimo is committed to consistently providing high quality products and services to you, our customers. We thank you for your patience and cooperation while we actively work to resolve this issue. . .
Quantity in commerce
42,908 units
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Africa, Asia, Australia/New Zealand, Europe, Canada, Latin America, and Middle East / SW Asia.
Product codes and lots
Catalog Number 9190 9193 9191 9192 9212 9213 9218 9219 9224 9282 9285 2883 9194

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1187-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1187-2013Class IITerminatedVoluntary: Firm initiated