Z-1099-2013Class II

Torrent" irrigation tubing

United States Endoscopy Group, Inc. / initiated 2013-03-07 / Terminated / root cause: design

US Endoscopy Group Inc began a recall of Torrent" irrigation tubing. Product Usage: The Torrent irrigation system (tubing and accessories to accommodate various endoscopes and irrigation pumps) is... on . FDA classified it as Class II and gives the reason as "The firm was notified by their customers that there was a water leakage from the white tubing connector". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1099-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
United States Endoscopy Group, Inc.
Product
Torrent" irrigation tubing. Product Usage: The Torrent irrigation system (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via irrigation fluids, such as sterile water, during gastrointestinal endoscopic procedures when used in conjunction with an irrigation pump or electrosurgical unit.
Product code
OCX Endoscopic Irrigation/Suction System
Reason for recall
The firm was notified by their customers that there was a water leakage from the white tubing connector.
Root cause tag
design
FDA root cause
Device Design
Action
US Endoscopy Group sent an Urgent Recall Notification letters to their customers. The letter identified the affected product, problem and actions to be taken. They instructed their customers to remove any of the affected products from their inventory and quarantine them. They also requested each customer complete the attached Customer Response Card and return it to US Endoscopy via fax at 1-440-639-4495 or by email at [email protected] as soon as possible. For questions contact Customer Service at 1-800-769-8229.
Quantity in commerce
21,189 units
Distribution
USA Nationwide Distribution - including the states of: AK, AL, AZ, CA, CO, CT, DE, FL, GA, ID, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI & WV.
Product codes and lots
Product code: 00711560; Lot #'s: 79687, 79688, 79689, 79690, 79691, 79692, 79693, 79694, 79695, 79696 80165, 80166, 80709, 80710, 80711, 80712, 80713, 80714, 80715, 80716, 80717, 80718, 80719. 80720, 80721, 80722, 80723, 81245, 81246, 81247, 81248, 81249, 81250, 81251, 81252, 81253, 81254, 81255, 81256, 81257, 81258, 81259, 81982, 81983, 81984, 81985, 81986, 81987, 81988, 81989, 81990, 81991, 82567, 82568, 82569, 82570, 82938, 82939, 82940, 82941, 82942, 82943, 82944, 82945, 82946, 82947, 82948, 82949, 82950, 82951, 82953, 82954, 82955, 82956, 84615, 84616, 84617, 84618, 84619, 84620, 84979, 84981, 85557, 86026, 86028, 86434, 86435 & 86436.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1099-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1099-2013Class IITerminatedVoluntary: Firm initiated