Z-1117-2013Class II

The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in...

Abbott Medical / initiated 2013-03-21 / Terminated / root cause: manufacturing process

Hospira Inc. began a recall of The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is... on . FDA classified it as Class II and gives the reason as "During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1117-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients.
Product code
FRN Pump, Infusion
Reason for recall
During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Hospira sent an Urgent Device Field Correction letter dated March 21, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to refrain from using the affected produc and contact Hospira at 1-800-441-4100 to arrange for return of their device for repair. Customers were directed to notify their customers if they further distributed the product and confirm they have done so by return of the reply form to Stericycle. For questions regarding this recall call 224-212-4892.
Quantity in commerce
1) List Number 13086: 286 pumps; 3) List Number 13087: 45,376 pumps; 4) List Number 13088: 23,492 pumps
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.
Product codes and lots
List Numbers: 13086, 13087, 13088

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1117-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1116-2013Class IITerminatedVoluntary: Firm initiated
Z-1117-2013Class IITerminatedVoluntary: Firm initiated