Z-1185-2013Class II

Medi-Trace Cadence Pediatric Radiotransparent Defibrillation Electrode Product ID: 22550P. Intended for use...

Medtronic, Inc. / initiated 2013-04-08 / Terminated / root cause: manufacturing process

Covidien LLC began a recall of Medi-Trace Cadence Pediatric Radiotransparent Defibrillation Electrode Product ID: 22550P. Intended for use in cardioversion, pacing, ECG monitoring, and... on . FDA classified it as Class II and gives the reason as "Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1185-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medi-Trace Cadence Pediatric Radiotransparent Defibrillation Electrode Product ID: 22550P. Intended for use in cardioversion, pacing, ECG monitoring, and defibrillation procedures.
Product code
DRO Pacemaker, Cardiac, External Transcutaneous (Non-Invasive)
Reason for recall
Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Covidien sent an "URGENT MEDICAL DEVICE RECALL" letter dated April 8, 2013 to all affected customers. The letter identifies the product, problem and requests that customers immediately stop using and segregate any affected product from inventory. Questions:Call 1-800-962-9888 option 8, then extension 2500.
Quantity in commerce
7485 (sets of 2)
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Korea, Japan, Canada, Australia, Brazil, Belgium, Singapore, and South Africa.
Product codes and lots
Lot Numbers: 228651, 232146, 235646X, 301833X, 303928X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1185-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1184-2013Class IITerminatedVoluntary: Firm initiated
Z-1185-2013Class IITerminatedVoluntary: Firm initiated